“ú–{Œê “ú–{Œê
F0101VR web-catalog (en) rev.1.01 20181121
 "Does your DMM measure it for sure?"


"Ultra-high-precision" that fit in your shirt pocket

2.5V±0.02%(max), 5kΩ±0.01%(max)
F0101VR Pocket Reference
Ultra-high-precision Voltage and Resistance Reference
with reference capacitor and resistive temperature detector



Make-to-order product 67,000JPY (Tax excl., Reference pricej

General description

F0101VR is a pocket sized, Ultra-High-Precision, Voltage and Resistance Reference, which is especially suitable for pre-operation inspection of digital multimeters of upto 4-1/2 digits precision. You can make one point calibratin of Voltage of 2.5V, Resistance of 5kĦ and Capacitance of 5nF, with confidence of the case-temperature that can be measured by the on-board resistive temerature detector. F0101VR can be delivered with an optional tag of the calibration data (Opt.T) of which is done at the factory delivery with calibrated equipments that have the traceability to the National standards.

Specifications

Warranty period 1 year (Opt.T)
Operating temperature range  +25±15Ž (0`80%R.H.)
Storage temperature range-20`+60Ž (0`80%R.H.)
Size W38.6 × H15.5 × D78.6 [mm]
Weight 40 [g]
Battery CR2032 × 1
Power dissipation ≃3mW (No load)
Terminals M3×3.5 Brass, Nickel base, gold plated. (Standard tightening torque 0.315 [N·m])
⚠️ If the tightening torque exceeds 0.5[N·m] , there is a risk of permanent damage of the circuit board.

Pilot lumps (LED)
Good Wthite Power good (Vbattery≥2.7V±1%)
Low Red Low battery, or over load
Voltatge (2.5V-terminal — 0V-terminalj
Nominal voltage2.5VTc=+25Ž
Tolerance±500μVTc=+25Ž
Calibration value accuracy±20ppmTc=Tcal, Vbattery=3V | Opt.T
Stabilization time 1s(max)Tc=+25Ž, Vo≤Von±100uV, RL≥1MΩ
Temperature coefficient±3ppm/Ž(max)0Ž≤Tc≤+60Ž
Long-term stability±20ppm/1000hours(typ)Tc=+25Ž
Maximum load current±1mATc=+25Ž
Load regulation±50μV/mA(max)Tc=+25Ž, Vbattery=3V
Line regulation±6μV(max)Tc=+25Ž, Vbattery=3V±10%
Noise voltage1.45μVP-P(typ)f = 0.1 to 10Hz
45nV/√Hz(typ)f = 1kHz
Resistance (5kΩ-termianl — 2.5V-terminalj
Nominal resistance5kΩTc=+25Ž
Tolerance±0.5ΩTc=+25Ž
Calibration value accuracy±50ppmTc=Tcal | Opt.T
Stabilization time 1s(max)Tc=+25Ž, R≤Rcal±10ppm
Temperature coefficient ±1.8ppm/Ž(max)0Ž≤Tc≤+60Ž
Maximum power dissipation 20mWΔR≤1ppm
Long-term stability50ppm/2000hours(typ)Tc=70Ž


Capacitance (5nF-terminal — 0V-termianl)
Nominal capacitance 5nFTc=+25Ž
Tolerance±80pFTc=+25Ž
Calibration value accuracy ±0.6%Tc=Tcal | Opt.T
Temperature coefficient +0|-30ppm/Ž(max)0Ž≤Tc≤+60Ž
Maximum voltage 25V(max)
Insulation resistance 10GΩ(min)
Resistive-Temperature-Detector (Rtd-terminal — 0V-terminal)
Tc
Ž
Rmin
Rtyp
Rmax
2012.02312.09912.176
2111.57111.64211.712
2211.13911.20411.269
2310.72510.78510.845
2410.32910.38410.439
259.950010.00010.050
269.58169.63239.6830
279.22889.28009.3313
288.89088.94258.9942
298.56708.61908.6711
308.25668.30898.3613
RTD= kΩ Tc °C

Abbreviations

ΔR:Variation of R
ΔV:Variation of Vo
C:Capacitance between 5nF-terminal and 0V-termianl
f:Frequency
R:Resistance between 5kΩ-termianl and 2.5V-terminal
Rcal:R value at the calibration
RL:Load resistance connected between 2.5V-terminal and 0V-terminal
Rtd:Resistive Temperature Detector
Tc:Case temperature
Tcal:Ambient temperature at the calibration
Vbattery:Battery voltage
Vcal:V at the calibrationon
Vo:Voltage between 2.5V-termianl and 0V-terminal when the Power switch is pressed
Von:Vo at no-load steady state

Utilizing as a current reference

Since F0101VR has ultra-high-precision voltage and resistance, it can be used for DC current calibration of full-scale range of 500μA ∼ 1mA of upto 4-1/2 digits DMM.
When the input resistance of the DMM at the DC current measurement Rs is known, measuring the DC current from the 5kΩ-terminal to the 0V-terminal, the measurement current, i.e. reference current Iref  is described as,
I_{ref} &=&  \frac{V_{ref}}{R_{ref} + R_s}
For example, in case of Rref = 5000±0.5Ω, Rs = 101±0.5Ω, Iref  will be,
I_{ref} &=&  \frac{2.5  \pm 0.0005}{5000 \pm 0.5 + 101 \pm 0.5} \\
\\
& \approx & 490.10  \rm{\mu A} \pm 0.04\%
Therefore, it can be used as the reference current of 490.10μA with ±0.04% accuracy.
Similarly, if Rs is less than 1Ω, it can be used as that of 499.95μA±0.04%.

Long-term stability

This is an example of long-term stability characteristics on a bench in an air-conditioned environment, which is considered to be common in the development of electronic circuits. After aging for 15 days, the variation during 672 hours at 20.5 to 27.0 °C was }3.5 ppm for 3 &sigma.
Note : The measured data includes missing time due to battery replacement or OS update.


Appendix — Examples of popular DMM and its accuracy, for electronic circuits

Maker Product CountsDCV
[V]
Accuracy
[%]
Resistance
[kΩ]
Accuracy
[%]
DCA
[μA]
Accuracy
[%]
Capacitance
[nF]
Accuracy
[%]
Comment
FinetuneF0101VRN.A.Vo=2.5 ±0.02Ro=5.0 ±0.01 (Io=500) ±0.04 Co=5.0 ±1.0
FinetuneF0101VN.A.Vo=2.5 ±0.02Ro=5.0 ±0.05 (Io=500) ±0.08 Co=5.0 ±1.0
Fluke 287 50000 5.0000 ±0.025 50.000 ±0.05 500.00 ±0.075 10.00 ±1.0
Fluke 87V 6000 6.000 ±0.05 6.000 ±0.2 600.0 ±0.2 10.00 ±1.0LT1790, 0.1 %, 25 ppm/°C
Fluke 83V 6000 6.000 ±0.1 6.000 ±0.4 600.0 ±0.4 10.00 ±1.0
HiokiDT4281
DT4282
60000 6.0000 ±0.025 6.0000 ±0.03 600.00 ±0.05 10.00 ±1.0
Hioki DT4253 6000 6.000 ±0.3 6.000 ±0.7 600.00±0.8(1.000μF)±1.9
KaiseKT-2011 50000 5.0000 ±0.03 50.000 ±0.1 500.00 ±0.15 50.00 ±0.8
KeysightU1252B
U1253B
50000 5.0000 ±0.025 50.000 ±0.05 500.00 ±0.05 11.000 ±1.0
KeysightU1251B 50000 5.0000 ±0.03 50.000 ±0.08 500.00 ±0.1 11.000 ±1.0
KeysightU1271A
U1272A
U1273A
U1273AX
33000 3.0000 ±0.05 30.000 ±0.2 3000.0 ±0.2 10.000 ±1.0
KeysightU1241C
U1242C
10000 10.000 ±0.09 10.000 ±0.2 1000.0 ±0.1 (1000.0) ±1.0
KeysightU1241B
U1242B
10000 10.000 ±0.09 10.000 ±0.3 1000.0 ±0.1 (1000.0) ±1.2
Keysight U1233A 6000 6.000 ±0.5 6.000 ±0.9 600.0 ±0.1 (1000) ±1.9
Sanwa PC7000 50000 5.0000 ±0.03 50.000 ±0.2 5000.0 ±0.6 50.00 ±0.8
SanwaPC720M
PC710
PC700
9999 9.999 ±0.06 6.000 ±0.1 600.0 ±0.2 60.00 ±0.8
Sanwa PC773 11000 11.000 ±0.28 11.000 ±0.3 1100.0 ±0.5 11.000 ±2.0
Sanwa CD732 6000 6.000 ±0.5 6.000 ±1.2 600.0 ±1.5 40.00 ±5.0
YokogawaTY720 50000 5.0000 ±0.025 50.000 ±0.05 500.00 ±0.2 50.00 ±1.0
YokogawaTY520 6000 6.000 ±0.09 6.000 ±0.4 600.0 ±0.2 10.00 ±2.0
YokogawaCA150 50000Vo=2.5 ±0.040Ro=5.0 ±0.05 Io=500 ±0.625 N.A. -
YokogawaCA310 50000Vo=2.5 ±0.035 N.A. - Io=500 ±0.615 N.A. -
⚠️ These values are excerpts from each company's DMM instruction manuals. Please refer to the specification of each DMM for accuracy.